| Filter Sterilization System |
过滤器灭菌系统 |
|
Power Supply |
电源 |
|
Preparatory Work |
准备 |
|
Ready state |
准备就绪 |
|
Sterilization |
灭菌 |
|
Alarm A-level |
A级报警 |
|
Alarm C-level |
C级报警 |
|
Starting Sterilize |
开始杀菌 |
|
|
|
|
Alarm Number |
报警序号列表 |
| 1 |
Vaporizer 1 air pressure low alarm |
雾化器1空气压力过低 |
| 2 |
Vaporizer 1 air pressure high alarm |
雾化器1空气压力过高 |
| 3 |
Vaporizer 2 air pressure low alarm |
雾化器2空气压力过低 |
| 4 |
Vaporizer 2 air pressure high alarm |
雾化器2空气压力过高 |
| 5 |
Air pressure switch 11 trouble |
空气压力开关11故障 |
| 6 |
Air pressure switch 21 trouble |
空气压力开关21故障 |
| 7 |
MV12 abnormal 1 |
MV12异常1 |
| 8 |
MV22 abnormal 1 |
MV22异常1 |
| 9 |
MV12 abnormal 2 |
MV12异常2 |
| 10 |
MV22 abnormal 2 |
MV22异常2 |
| 11 |
MV12 abnormal 3 |
MV12异常3 |
| 12 |
MV22 abnormal 3 |
MV22异常3 |
| 13 |
Vaporizer 1 inlet low temperature |
雾化器1入口温度过低 |
| 14 |
Vaporizer 1 inlet high temperature |
雾化器1入口温度过高 |
| 15 |
Vaporizer 2 inlet low temperature |
雾化器2入口温度过低 |
| 16 |
Vaporizer 2 inlet high temperature |
雾化器2入口温度过高 |
| 17 |
Vaporizer 1 chemical liquid after jet outlet low temperature |
雾化器1药剂喷出温度过低 |
| 18 |
Reserve |
保留 |
| 19 |
Vaporizer 2 chemical liquid after jet outlet low temperature |
雾化器2药剂喷出温度过低 |
| 20 |
Reserve |
保留 |
| 21 |
Supply air pressure low alarm |
空气供给压力过低 |
| 22 |
PLC abnormal |
PLC异常 |
| 23 |
FFU1.2 alarm |
FFU1、2报警 |
| 24 |
Vaporizer 1 air pressure switch12 alarm |
雾化器1空气压力开关12报警 |
| 25 |
Vaporizer 2 air pressure switch22 alarm |
雾化器2空气压力开关22报警 |
| 26 |
Chemical liquid pressure switch1A alarm |
药剂压力开关1A报警 |
| 27 |
MV1A abnormal 1 |
MV1A异常1 |
| 28 |
MV1A abnormal 2 |
MV1A异常2 |
| 29 |
MV1A abnormal 3 |
MV1A异常3 |
| 30 |
Chemical liquid tank bottom |
药剂罐底(补给药剂) |
| 31 |
Chemical liquid pressure switch 13 alarm |
药剂压力开关13报警 |
| 32 |
MV11 abnormal 1 |
MV11异常1 |
| 33 |
MV11 abnormal 2 |
MV11异常2 |
| 34 |
MV11 abnormal 3 |
MV11异常3 |
| 35 |
MV11 open timelimit |
MV11打开时限 |
| 36 |
Chemical liquid pressure switch23 alarm |
药剂压力开关23报警 |
| 37 |
MV21 abnormal 1 |
MV21异常1 |
| 38 |
MV21 abnormal 2 |
MV21异常2 |
| 39 |
MV21 abnormal 3 |
MV21异常3 |
| 40 |
MV21 open timelimit |
MV21打开时限 |
| 41 |
loadcell controller 1 abnormal |
重量控制传感器1异常 |
| 42 |
loadcell 1 residual quantity alarm |
称重传感器1剩余量报警 |
| 43 |
loadcell controller 2 abnormal |
重量控制传感器2异常 |
| 44 |
loadcell 2 residual quantity alarm |
称重传感器2剩余量报警 |
| 45 |
chemical liquid pressure switch 1B alarm |
药剂压力开关1B报警 |
| 46 |
MV1B abnormal 1 |
MV1B异常1 |
| 47 |
MV1B abnormal 2 |
MV1B异常2 |
| 48 |
MV1B abnormal 3 |
MV1B异常3 |